GRETINA LN filling system

Status (Feb 2016)

LN supply ON (valve 2 in wall manifold) is c0.2/relay #1, i.e. open main supply to GRETINA with opv 1

NOTE: don't use 'cooldown' for manifold 0 (wall manifold). If bypass valve c0.3 is opened LN from the Dewar actually will flow back through valve c0.1 (even closed) into vent line, as Dewar pressure is higher the building supply pressure.

manifold ovl Pt100 ovl Pt100 #adc valve valve #relay
0 n.a. (back pressure) c0.3 2
1 c32.1 182 c12.1 40
2 c32.1 182 c12.1 40
3 c33.1 183 c14.1 56
4 c33.1 183 c14.1 56
detector LN ID ovl Pt100 ovl Pt100 #adc valve valve #relay manifold
Dewar 0 c30.1 180 c0.4 3 0
Q1 1 c1.2 7 c12.3 42 1
Q2 2 c2.2 13 c11.2 33 2
Q3 3 c3.2 19 c14.4 59 3
Q4 4 c4.2 25 c12.5 44 1
Q5 5 c5.2 31 c14.2 57 4
Q6 6 c6.2 37 c12.2 41 1
Q7 7 c7.2 43 c14.5 60 3
Q8 8 c8.2 49 c11.4 35 2
Q9 9 c9.2 55 c12.6 46 2
Q10 10 c0.2 1 c14.3 58 4
Wall Box description valve valve #relay
1 building LN c0.1 0
2 to Gretina c0.2 1
3 DO NOT USE c0.3 2
4 exhaust c0.4 3

filling buffer Dewar: opv 0 & opv 3

NEVER open wall box valve 3 (opv 2), LN from buffer Dewar is then released into vault via exhaust!

Connector-Channel map LNmap

Misc

part numbers

  • LN sensor (new style): http://www.omega.com/ppt/pptsc.asp?ref=PR-10&Nav=hear01, 1/8 DIAMETER RTD PROBES, P# PR-10-2-100-1/8-2.5“-E, NOTE: 2.5” is custom made
  • Tube to Pipe fitting for PT100 (Swagelok # B-200-1-4BT; 1/8 Tube X 1/4 NPT Bored thru), NOTE: thread for needs to be modified. photo of modified fitting: fittingpt100.jpg
  • Replacement Springs from Lee Spring Company :P# LP 020E 04 S316 the correct fit that will only allow 1-3 PSI max.
  • Valve for detector fill line ASCO Red-Hat II 8263G209LT
  • ASCO valve in main manifold box is 8222G002LT 24VDC
  • Repair kit for manifold valve is 306399 Repair Kit, see quote asco_kit.xlsx

Maintenance